Close icon Close icon

FODL Multi-channel Field Datalogger

HIGH RELIABILITY

16 OR 32 CHANNELS

FIBER OPTIC DATALOGGER

The FODL-1600/3200 is a universal multi-channel fiber optic field datalogger ideally suited for performing multi-point temperature, pressure, strain and displacement measurements.

Description

The FODL-1600/3200 datalogger is designed for applications that require continuous monitoring of a large number of measuring points. The FODL-1600/3200 datalogger is capable of measuring the absolute cavity length of Fabry-Perot fiber optic transducers with great accuracy, providing highly reliable measurements.

It is compatible with fiber optic transducers, including temperature, strain, pressure, displacement and force and load. Roctest’s transducers feature complete immunity to microwave and RF radiation.

The FODL-1600/3200 has a 0.01% F.S. resolution (without averaging) and 0.025% F.S. precision. Use of a Flash ROM allows the user to easily upgrade the signal conditioner firmware. The FODL-1600/3200 has a non volatile memory buffer that can store up to 50 000 data samples and datalogging sequences. Duration and other operational parameters are easily programmable using RS-232 remote control and user-friendly.

The FODL-1600/3200 datalogger comes standard in a NEMA-4 enclosure that accommodates 16 to 32 channels. The RS-232 I/O port is used for control and data downloads.

Specifications
  

Accuracy

0.025% of F.S.

Resolution

0.01% of F.S.

Number of channels

16 or 32

Sampling rate

20 Hz

Switching time

150 ms (in scan mode: time to switch between two channels)

Averaging

1 to 500 samples

Display

None

Operating mode

RS-232 (software included)

Datalogging

50 000 samples; programmable datalogger

Analog outputs

None

Communication

RS-232

Diagnostic

Yes

Upgradeability – firmware

Flash ROM upgradeable

Upgradeability – channels

No

Light life expectancy

~40 000 hours of continuous use (MTBF)

Fiberglass enclosure

NEMA-4 enclosure (other enclosures available)

System power requirements

 

  Voltage

9 to 15 V (AC/DC adapter included)

  Minimum current capacity

500 mA

  Consumption

5 W

Operating temperature

-20 to +40°C

Enclosure dimensions

445 x 495 x 225 mm (w x d x h)

Weight

17 kg

Ordering Info

Please specify:

  • Come as a complete unit
  • Transducers must be ordered separately